AvCount Lube Particle Counter – SA1900-0
The AvCount Lube particle counter determines particle concentration in oils with a viscosity of up to 200 mm2/s. FFKM seals have been fitted to resist aggressive oils.
• Simple operation
• Ideal for handling lubricating and hydraulic oils
• 3-minute test time
• Programmable via PC
With a simple test procedure and user-friendly features, operator skill or expertise is kept to a minimum.
AvCount Lube can hold up to 3 test methods with custom methods created on a PC before being uploaded using the supplied ProTrend control and analysis software.
Test progress and results for up to 6 size bands are displayed as cumulative particles/ml and ISO 4406, NAS 1638, and AS4059 cleanliness codes.
Supplied with: sample probe, waste pipe, universal voltage power supplies, USB cable, calibration certificate, and USB memory stick containing handbooks and ProTrend software for test method programming and data download.
Methods: ASTM D7619, ASTM D7647, IP 565, GJB 420-1987/420-A-1996/420B-2006, IEC 60970, ISO 4406, ISO 11500, ISO 60970, JIS B 9932:2012, JIS B 9933:2021, JIS B 9934:2012
Features
- Calibrated to ISO 11171, traceable to NIST SRM 2806d
- Supplied with ASTM D7647, ISO 60970, and ISO 4406 loaded as standard
- FFKM seals compatible with synthetic oils
- Suitable for fluids up to 200mm2/s viscosity
- Precise and reliable results
- Ultra-compact rugged, steel case suitable for portable use
- Data storage for up to 600 measurements
- Internal Double Pump System
- USB connection for PC connectivity
Integral Compressor
AvCount Lube comprises a light extinction automatic particle counter and a sample delivery unit. Designed to handle samples with a viscosity up to 200 mm2/s, the AvCount Lube has an integral compressor that pressurizes the sample chamber to 3 bar.
Calibration and verification:
AvCount Lite utilizes 16-point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated ‘Master’ AvCount Lite. Calibration can be carried out at a user site or at our authorized regional laboratories. Seta authorized training is required to perform a calibration. Verification can be performed using AvCount Verification Material.
ProTrend (SA1810-0) is supplied with the instrument and can report results to ISO 4406, NAS 1638, and SAE AS4059 (tables 1 and 2) cleanliness codes. Using the software, you can program other test methods and cleanliness codes with up to 16 size bands.
PARTICLE SIZE RANGE | ISO 11171: 4µm(b) to >70µm(b) |
---|---|
NUMBER OF MEASURING CHANNELS | 15 when connected to a PC, 6 when used as a stand-alone |
COUNTS PER MEASUREMENT (MAX) | 16.000.000 per 10 ml |
SAMPLE VISCOSITY (MAX) | 200 mm2/s |
SAMPLE TEMPERATURE RANGE | 0 to 70 °C |
RESULTS | 600 internal, unlimited on PC |
PUMP VOLUME & FLOW RATE | 10 ml aliquots @ 30 ml/min or timed |
TOTAL SAMPLE VOLUME USED (TYP) | 80 ml (includes rinse cycles) for ASTM D7647, 20 ml for other methods |
VOLTAGE | Automatic particle counter: 12 Vdc, 2.5A (mains adaptor supplied 100 to 240 Vac, 50/60 Hz, 650mA) Pressure delivery unit: 24 Vdc, 2A (mains adaptor supplied 100 to 240 Vac, 50/60 Hz, 1A) |
SIZE (HXWXD) | 50 x 32 x 28 cm |
WEIGHT | 16 kg |
CCCN CODE | Tariff 90318020 |