Setaflash Series 3 Active Cool, 33200-2

The Setaflash Series 3 ActiveCool Small Scale Flash Point Tester is the latest instrument in the Setaflash range. It has been designed to rapidly and efficiently carry out Flash / No Flash tests to determine the flash point of liquids and semi-solids in the 10 to 130°C (50 to 266°F) temperature range.

Methods: ASTM D3278, ASTM D3828, ASTM D7236, ASTM E502, BS EN ISO 3679, BS 2000-523, Classification of Dangerous Goods for Carriage: Viscous and Non-viscous Liquids, CLP Regulations EC No 1272/2008, EPA 1020 A & B, IP 523, IP 524 (Obs), IP 534, ISO 3680 (Obs), UN Class 3 Non-viscous Flammable Liquids

When used in Ramp mode the temperature of the sample increases at a rate of 2°C/min. An audible prompt to carry out a flash test occurs every 1°C rise up to 100°C (every 2°C rise above 100°C), until a flash is detected, allowing rapid determination of a flash point in accordance with ASTM D7236 and IP 534. The sample cup is heated / cooled by peltier cells. This gives very accurate and stable sub-ambient test temperatures and the benefit of rapid temperature normalisation between tests.

All functions on the Series 3 flash point tester are accessed via a two button keypad and the Seta Multifunctional And Rotational Test (SMART) control feature. The digital display shows test parameters, instrument status, and the test result. An audible prompt is sounded when an action is required from the operator. Gas for the test jet/shutter assembly is supplied from an integral gas tank, which is filled using a standard butane (lighter) refill cartridge.

The automatic flash detection system is tuned to ignore halo and other false flash effects, reducing the chance of mis-interpretation of the test result, and enhancing repeatability. Barometric pressure is measured by an internal barometer, and the correction is automatically applied to the test result.

The instrument is lightweight and suitable for mobile use when operated from the optional DC power adaptor.

Supplied with: 2ml syringe, mains lead and instruction manual.

  • Flash / No Flash & Ramp modes
  • 10 to 130°C temperature range
  • Flash point test in less than 2 minutes
  • Small sample size (2 or 4ml)
  • Automatic flash detection
  • Electronic peltier cooling
  • Audible and on-screen prompts
  • Portable operation
  • °C or °F display
Temperature Range 10 to 130°C (50 to 266°F)
Cup Material Aluminium
Test Modes Rapid Equilibrium and Ramp
Sample Size 2ml for flash points up to 100°C 4ml for flash points above 100°C
Test Duration, Rapid Equilibrium Mode 1 minute below 100°C, 2 minutes above 100°C, or user determined 1 to 99 minutes
Ramp Rate 2°C/min ramp
Heating and Cooling System Peltier Cell
Power 200W (maximum)
Voltage 100/240V, 50/60Hz (switchable) 12Vdc with optional power adaptor
Gas Supply Laboratory gas, 3kPa (0.44psi) maximum pressure or Butane from optional Portable Gas Tank 13667-0
Size (HxWxD) 21 x 28 x 14cm
Weight 3 kg
CCCN Code: Tariff 90268020

Lazar Scientific is your reliable source for Flash Point Testing Equipment.