574-271-7020 support@lazarsci.com

Sindie® On Line

Sindie Online is an industrial-grade process analyzer delivering continuous sulfur detection to monitor fuel streams and help to prevent tank contamination. It has many advantages over competing technologies: Powered by MWDXRF, Sindie Online has an exceptional signal-to-noise ratio using monochromatic excitation of the X-ray source characteristic line – a technique that is faster, safer and cleaner than UVF.

On-Line Sulfur Analysis in Petroleum Fuels

The Sindie® On-Line Analyzer is an industrial grade process sulfur analyzer with break-through detection capability for monitoring fuel streams as exacting as ultra low sulfur diesel and gasoline. This process analyzer presents the ultimate solution for pipeline terminals, where measurement speed and reliability are essential. The Sindie On-Line Analyzer is pow-ered by Monochromatic Wavelength Dispersive X-Ray Fluorescence (MWD XRF) technology, offering a limit of detection (LOD) of 0.6 ppm and a dynamic range of 3000 ppm. This direct and non-destructive measurement technique does not require sample conver-sion or consumable gasses and does not involve high temperature operations. The result: a robust process analyzer with minimal maintenance and unprecedented detection capability and measurement speed.

Application Areas:

  • Refinery: hydrotreating, hydrofiner, and blending processes
  • Refi nery: hydrotreating, hydrofiner, and blending processes
  • Pipeline terminals: interface cuts, custody transfer acceptance, and tank contamination prevention
  • Excellent detection capability: LOD: 0.6 ppm
  • Excellent detection capability: LOD: 0.6 ppm
  • Dynamic range from 0.6 ppm – 3000 ppm sulfur in diesel, gasoline, naptha, and kerosene
  • Continuous monitoring with programmable response times:
    • 30 seconds: pipeline interface cuts
    • 5 minutes: most demanding refi nery processes
  • Direct measurement without sample conversion; Analysis in ppm (wt)
  • No density conversion needed
  • No consumables, no gasses, no high temperature processes
  • Extremely low maintenance:
    • No heating elements
    • No quartz tubing
    • No columns
  • Dynamic window module design is operator independent and ensures measurement stability
  • Robust industrial design: wall mounted or stand alone
  • Outstanding linearity: One calibration curve for diesel and gasoline matrixes, over full dynamic range
  • Total sulfur determination using MWD XRF
  • Extended range available for measurements above 3000 ppm up to weight percent levels
Electrical
Input voltage: 115/230 VAC
Input frequency: 57-63 Hz
Requires 15 amp, dedicated circuit with disconnect
Entry Hub for 1” NPT conduit (RMC or IMC)
Purge
Instrument Air
< 0.5 scfm @ 40-115 psig
4 scfm during initial purge cycle (21 minutes @ 40 -115 psig)
Connections: (1) 3/8” SST tubing
Sampling Management System Included
Inlet: 1.5-3.0 gph @ 25-80 psig
Outlet: 1.5-3.0 gph @ 10-65 psig (Approximately 15 psig pressure drop)
Ambient Temperature Requirements
0-35o C (32-95o F)
Size and Weight
38”w x 18”d x 60”h
280 lbs.
Communication
(2) 4-20mA proportional to sulfur concentration (standard)
Modbus TCP (optional)
Modbus RS-485 (optional)
Modbus RS-232 (optional)
Ethernet (optional)
Sample Conditioning Systems
Optional: Particle and H2O Removal
Classification
Class I, Division 2, Groups B, C, D, T4A (Z-Purge)

Lazar Scientific is your reliable source for Elemental Analysis.