AvCount Lube Low-Visc Particle Counter – SA1910-0

0-0

The AvCount Lube Low-Visc particle counter determines particle concentration in oils with a viscosity of up to 64 mm2/s. FFKM seals have been fitted to resist aggressive oils.

• Simple operation
• Ideal for handling lubricating and hydraulic oils
• 3 minute test time
• Programmable via PC

With a simple test procedure and user-friendly features, operator skill or expertise is kept to a minimum.

AvCount Lube Low Visc can hold up to 3 test methods with custom methods created on a PC before being uploaded using the supplied ProTrend control and analysis software.

Test progress and results for up to 6 size bands are displayed as cumulative particles/ml and ISO 4406, NAS 1638, AS4059 cleanliness codes.

Methods:ASTM D7619, ASTM D7647

IP 565, GJB 420-1987/420-A-1996/420B-2006,, ISO 4406, ISO 60970

Features

  • Calibrated to ISO 11171, traceable to NIST SRM 2806b
  • Supplied with ASTM D7647, ISO 60970 and ISO 4406 loaded as standard
  • FFKM seals compatible with synthetic oils
  • Suitable for fluids up to 64 mm2/s viscosity
  • Precise and reliable results
  • Ultra compact rugged, steel case suitable for portable use
  • Data storage for up to 600 measurements
  • Internal Double Pump System
  • USB connection for PC connectivity

Calibration & Verification
AvCount Lube Low-Visc utilizes 16 point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated Master AvCount Lube Low-Visc. Calibration can be carried out at a user site or at our authorized regional laboratories.

ProTrend (SA1810-0) can report results to ISO 4406, NAS 1638 and SAE AS4059 (tables 1 and 2) cleanliness codes. Using the software, you can program other test methods and cleanliness codes with up to 16 size bands.

PARTICLE SIZE RANGE 4 – 100 µm(c) (in accordance with ISO 11171)
NUMBER OF SIZE BANDS Up to 16 when connected to Protrend, 6 when used as a stand-alone
MAXIMUM PARTICLE COUNT PER ML 60 000 particles/ml
SAMPLE VISCOSITY (MAX) 64 mm2/s
SAMPLE TEMPERATURE RANGE 0 – 70 °C
RESULTS 600 internal, unlimited on PC
SAMPLE FLOW RATE 30 ml/min ±5 ml/min
TOTAL SAMPLE VOLUME USED 80 ml (includes rinse cycles) for ASTM D7647, 20 ml other methods
POWER SUPPLY 100 – 240 Vac, 50/60 Hz
SIZE (HXWXD) 27 × 32 × 28 mm
WEIGHT 7 kg

Lazar Scientific is your reliable source for Fuel Cleanliness Testing Equipment.

CFPP

Cloud Point and Pour Point

Color Measurement

Corrosion

Density Meters

Distillation

Elemental Analysis

Flash Point Testers

Foaming, Air Release, Emulsification

Freezing Point

Fuel Cleanliness

Oxidation Stability

Penetrometers / Grease Workers

Polarimeters

Refractometers

Inline Process Refractometers

Tribology

Vapor Pressure Testers

Viscometers

Miscellaneous Lab Apparatus