AvCount Lube Low-Visc Particle Counter – SA1910-0

0-0

The AvCount Lube Low-Visc particle counter determines particle concentration in oils with a viscosity of up to 64 mm2/s. FFKM seals have been fitted to resist aggressive oils.

• Simple operation
• Ideal for handling lubricating and hydraulic oils
• 3 minute test time
• Programmable via PC

With a simple test procedure and user-friendly features, operator skill or expertise is kept to a minimum.

AvCount Lube Low Visc can hold up to 3 test methods with custom methods created on a PC before being uploaded using the supplied ProTrend control and analysis software.

Test progress and results for up to 6 size bands are displayed as cumulative particles/ml and ISO 4406, NAS 1638, AS4059 cleanliness codes.

Methods:ASTM D7619, ASTM D7647

IP 565, GJB 420-1987/420-A-1996/420B-2006,, ISO 4406, ISO 60970

Features

  • Calibrated to ISO 11171, traceable to NIST SRM 2806b
  • Supplied with ASTM D7647, ISO 60970 and ISO 4406 loaded as standard
  • FFKM seals compatible with synthetic oils
  • Suitable for fluids up to 64 mm2/s viscosity
  • Precise and reliable results
  • Ultra compact rugged, steel case suitable for portable use
  • Data storage for up to 600 measurements
  • Internal Double Pump System
  • USB connection for PC connectivity

Calibration & Verification
AvCount Lube Low-Visc utilizes 16 point calibration. Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated Master AvCount Lube Low-Visc. Calibration can be carried out at a user site or at our authorized regional laboratories.

ProTrend (SA1810-0) can report results to ISO 4406, NAS 1638 and SAE AS4059 (tables 1 and 2) cleanliness codes. Using the software, you can program other test methods and cleanliness codes with up to 16 size bands.

PARTICLE SIZE RANGE 4 – 100 µm(c) (in accordance with ISO 11171)
NUMBER OF SIZE BANDS Up to 16 when connected to Protrend, 6 when used as a stand-alone
MAXIMUM PARTICLE COUNT PER ML 60 000 particles/ml
SAMPLE VISCOSITY (MAX) 64 mm2/s
SAMPLE TEMPERATURE RANGE 0 – 70 °C
RESULTS 600 internal, unlimited on PC
SAMPLE FLOW RATE 30 ml/min ±5 ml/min
TOTAL SAMPLE VOLUME USED 80 ml (includes rinse cycles) for ASTM D7647, 20 ml other methods
POWER SUPPLY 100 – 240 Vac, 50/60 Hz
SIZE (HXWXD) 27 × 32 × 28 mm
WEIGHT 7 kg

Lazar Scientific is your reliable source for Fuel Cleanliness Testing Equipment.