The Clora ® compact analyzer measures total chlorine in hydrocarbons such as aromatics, distillates, heavy fuels, crude oils, and water.

This state-of-the-art technology complies with ASTM D7536 and D4929C and delivers unparalleled accuracy and precision for petroleum and petrochemical applications where simple, quick, and reliable analysis is critical.

Methods: ASTM D7536, ASTM D4929C

Chlorine Analysis in Liquid Hydrocarbon and Aqueous Solutions

This model utilizes standard XRF sample cups

  • Easy to use
  • Extremely low maintenance
  • Total chlorine analysis
  • For use in multiple application areas
  • Hassle-free sample preparation

Easy to use
LIMS integration for data management and transfer, intuitive 10-inch touch screen, measurement time 10-999 s, barcode reader auto fills sample name, USB connectivity, storage capacity for more than 50,000 results, supports up to 30 calibration curves

Extremely low maintenance
No gasses, columns, heating elements, or quartz tubing.

Total chlorine analysis
Petroleum products, biofuels, aromatics and other chemicals, and water.

For use in multiple application areas
Refineries, petrochemical and additive plants, pipeline terminals, and test laboratories.

Hassle-free sample preparation
Utilizes Accucell pre-assembled and pre-vented sample cups for enhanced precision, extreme ease-of-use, and enhanced productivity.

Accu-flow: No
Calibration curves: Up to 30 calibration curves. Automatic and Manual Calibration functionality.
Catalysts: No
Communication Capabilities: USB, Ethernet: LIMS Connectivity
Consumable/Utility Gas: Internal Vacuum Pump
Dimensions: Depth: 21 in / 54 cm
Dimensions: Height: 16.5 in / 42 cm
Dimensions: Width: 15.8 in / 40 cm
Element: Chlorine / 0.13 ppm
Measurement Time: User selectable 10 – 999 s, 300 – 600 s typical
Method Name: ASTM D7536 and D4929C
Options: USB Printer (Receipt or Full Page), Consumables, Calibration Standards, Spare Parts.
Range: 0.13 ppm – 4%

Repeatability:
ppm – r

Xylene
0.05 – 0.15
1 – 0.17
5 – 0.31
10 – 0.50

Crude Oil
5 – 0.4
10 – 0.8
50 – 1.2

Water
5 – 0.6
10 – 1.0
50 – 1.5

Reproducibility:
ppm – R

Xylene
1 – 0.29
5 – 0.53
10 – 0.90

Crude Oil
5 – 0.8
10 – 1.4
50 – 2.4

Water
5 – 1.0
10 – 1.5
50 – 3.2

Sample Cup Type: Accucell
Sample Matrix: Hydrocarbon/ water
Specific Technology: MWDXRF
Type: Benchtop
Weight: 75 lbs / 34 kg
What’s included?: Quick-Start Guide, Operations Manual, radiation survey, power cord, 6′ drain tubing, spare fan filter, spare fuse, 100 Accucell Sample Cups, sample stand, spare primary window, and one calibration set; 0-500 ppm Cl in Mineral Oil (6 x 10 mL), bar code reader. Printer optional.
XRF Film Type: Etnom