Clora with Autosampler – 401430-01M

Clora with Autosampler – 401430-01M

Chlorine Analysis in Liquid Hydrocarbons, Aqueous Solutions, and Catalyst

Clora benchtops are compliant with ASTM D7536 and D4929 standards and deliver a limit of detection of 0.13 ppm and a dynamic range of up to 3000 ppm. Clora is powered by MWDXRF, the same technology found in our signature sulfur analyzer, Sindie. This direct measurement technique does not require gasses or high-temperature processes equating to easy operation and minimum maintenance requirements.

Methods: ASTM D7536, D4929

Features

This model features an 8-cell autosampler and utilizes Accucell sample cups for hassle-free sample prep.

  • Easy to use
  • Extremely low maintenance
  • Total chlorine analysis
  • For use in multiple application areas
  • Hassle-free sample prep

Easy to use
User-friendly with a touch screen interface. Just plug-it-in and measure results with one touch: power is only utility.

Extremely low maintenance
No conversion gasses, columns, heating elements, or quartz tubing

Total chlorine analysis
Aqueous solutions and aromatic products to heavy fuels, crudes, and catalyst.

For use in multiple application areas
Refinery labs, pipeline terminals, additive plants, testing vans, and inspection laboratories

Hassle-free sample prep
Utilizes Accucell pre-assembled and pre-vented sample cups for enhanced precision, extreme ease-of-use, and enhanced productivity.

Accu-flow: No
Calibration curves: 8 calibration curves. Automatic and Manual Calibration functionality.
Catalysts: No
Communication Capabilities: Printer, Ethernet: LIMS Connectivity (optional)
Concentration: In ppm: 5, 10, 50 (crude oil)
Consumable/Utility Gas: Vacuum
Dimensions: Depth: 19.7 in / 50.0 cm
Dimensions: Height: 13.4 in / 34.0 cm
Dimensions: Width: 14.6 in / 37.0 cm
Element: Chlorine / 0.13 ppm
Measurement Time: User selectable 30-900 s, 300-600 s typical
Method Name: ASTM D7536, ASTM D4929
Options: LIMS Connectivity
Range: 0.13 ppm – 3000 ppm

Repeatabilityppm – r

Xylene
1 – 0.17
5 – 0.31
10 – 0.50

Crude Oil
5 – 0.4
10 – 0.8
50 – 1.2

Water
5 – 0.6
10 – 1.0
50 – 1.6

Reproducibility: ppm – R

Xylene
1 – 0.29
5 – 0.53
10 – 0.90

Crude Oil
5 – 0.8
10 – 1.4
50 – 2.4

Water
5 – 1.0
10 – 1.5
50 – 3.2

Sample Cup Type: Accucell
Sample Matrix: hydrocarbon
Specific Technology: MWDXRF
Type: Benchtop
Weight: 70 lbs / 32 kg
What’s included?: Quick-Start Guide, Operations Manual, radiation survey, power cord, 6′ drain tubing, spare fan filter, printer, printer cable, printer ink, 1 roll of printer paper, spare fuse, 100 Accucells, 100 sample films, sample stand, and 1 calibration set; 0-500 ppm Cl in Mineral Oil – 6 x 10 mL.
XRF Film Type: Etnom