Seta Oil Test Centrifuge – 6 place – 90100-0
The Seta Oil Test Centrifuge is a heated chamber centrifuge used to determine water and sediment in oils. The large diameter chamber, with six place rotor and cushioned adaptors, allows the use of 6 and 8-inch conical, pear-shaped, trace sediment and 12.5 ml tubes.
The swing-out rotor ensures that samples are held near vertical when at rest, allowing precise measurement of sediment. The centrifuge features a hinged lid that opens to the fully upright position, allowing easy loading and unloading of tubes between tests.
Speed, temperature, and timing are microprocessor controlled. Run parameters are entered via a liquid-proof ‘touch panel’ and digital display.
Methods: ASTM D91, ASTM D893, ASTM D1290, ASTM D1796, ASTM D2273, ASTM D2709, ASTM D2711, ASTM D4007, ISO 3734, ISO 9030
Features
- Isothermal heaters fitted internally to the centrifuge chamber
- Temperature control using a unique algorithm to compensate for air resistance heating
- 6 place swing out rotor
- Static, near vertical bucket positioning
- Specifically designed for testing oils
- Heated chamber, ambient to 80 °C
- Pre-heat facility to ensure the bowl is at test temperature prior to the test
- Universal buckets for 12.5 ml, 6 and 8-inch glassware
- Microprocessor control with 99 memories
- Optional trolley
Precise heating for accurate results
Isothermal heaters fitted internally to the centrifuge chamber give precise temperature control to within +/-1 °C.
Fast and easy operation
9 program memories allow a user to configure the instrument to automatically operate to the specified test parameters, saving time, easing operation and assisting repeatability. A ‘Quick Run’ mode allows short runs of up to 99 seconds.
Safety features
A motor-driven lid latch with a safety interlock prevents the centrifuge operation while the lid is unlocked. Imbalanced detection and over-temperature protections protect the motor and chamber from damage.
Setafuge Hazard Reduction System
The optional Setafuge Hazard Reduction System (90010-0) can be fitted into the centrifuge to extract vapors from the chamber and significantly reduce vapor hazards.
Supplied with;
Rotor
Buckets and cushions
MAXIMUM RELATIVE CENTRIFUGAL FORCE (RCF) | 2210 g |
---|---|
MAXIMUM ROTATIONAL SPEED | 3000 rpm |
TEMPERATURE RANGE | Ambient +5 °C in the range of 20 to 80 °C (1 °C steps) |
TIMER RANGE | 10 sec to 100 hours or continuous operation |
CONTROL | Microprocessor with 99 memory capacity |
CAPACITY | 6 Samples |
VOLTAGE RANGE | 230 Vac, 50/60 Hz (Optional transformer for 110/120 V) |
POWER | 950 W |
SIZE (HXWXD) | 410 × 560 × 640 mm |
WEIGHT | 72.5 kg |
CCCN CODE | Tariff 84211970 |
Accessories, Spares, Calibration, and Training
-
Tube 8 inch x 100 ml cone-shaped (Pack of 6) – 90040-0
-
Bucket Adaptor for 90040-0 & 90060-0 (pack of 4) – 90001-2
-
Tube 6 inch x 100 ml pear-shaped (Pack of 6) – 90050-0
-
Tube 6 inch x 100 ml cone-shaped (pack of 6) – 90011-0
-
Transformer for 120 V 60 Hz Operation – 90016-0
-
High Level Centrifuge Trolley – 90015-0
-
Low Level Centrifuge Trolley – 90014-0
-
Bucket Adaptor for 90011-0 Tube (pack of 4) – 90005-2
-
Bucket Adaptor for 90050-0 (Pack of 4) – 90003-2
-
Stoppers (Pack of 50) – 90041-2
-
Centrifuge_Tube_90012-0
-
Adaptor for 90012-0 Centrifuge Tube (Pack of 4) – 90002-0
-
Setafuge Hazard Reduction System – 90010-0
-
Bottom Cushions (pack of 2) – 90000-301
-
Digital Photo Tachometer – 99960-2
-
Thermometer Digital: -199 to 199 °C UKAS Calibrated – 82012-0
-
Rotor Collet – 90000-344