Seta Semi-Automatic Tag Open Cup Flash Point Tester – 13230-3

The instrument consists of a stainless steel case in which is mounted the electrical heater, liquid bath and glass test cup. The button-operated, electrically driven sweep arm can use either mains gas or LPG from the optional Portable Gas Tank (13667-0). At the rear of the raised panel is a gas inlet block, comprising a 10 mm diameter nozzle and a pinch valve for adjusting the size of the test flame. A test flame size bead is mounted on the top of the liquid bath.

Methods: ASTM D1310, ASTM D3143

Features

  • Ambient to 93 °C
  • Variable temperature rise rate
  • 500 ml Bath
  • Electrically actuated ignition taper

A leveling device, required by the method for checking the position of the ignition taper and the sample level in the glass cup, is supplied.

Supplied with: instruction manual and mains lead.

CUP TYPE Open Cup
TEMPERATURE RANGE Ambient to 93 °C
RAMP RATE Variable, manually controlled
BATH FLUID 500 ml Water or Silicone Fluid
POWER 500 W maximum
VOLTAGE 220/240 V, 50/60 Hz
SIZE (HXWXD) 33 x 31 x 29 cm
WEIGHT 6 kg
CCCN CODE Tariff 90268020