Sindie 2622 Gen2 with Autosampler - Accu-cell (vacuum) - 402169-01M

Sindie 2622 Gen2 with Autosampler – Accu-cell (vacuum) – 402169-01M

Sulfur Analysis with Compliance Flexibility

Sindie 2622 complies with ASTM D2622, D7039, and ISO 20884 methods, enabling complete flexibility in sulfur analysis. No compromises in detection, performance, and reliability — it is the ideal sulfur analytic solution from ultra-low sulfur diesel and gasoline, to heavy fuel oil and crude.

Methods: ASTM D2622, D7039, ISO 20884

Features

This model features an 8-cell autosampler and with Accu-cell sample cups for hassle-free sample prep.

  • Easy to use
  • Extremely low maintenance
  • Total sulfur analysis
  • For use in multiple application areas
  • Hassle-free sample prep

Easy to use
User-friendly with a touch screen interface. Just plug-it-in and measure results with one touch: power is only utility.

Extremely low maintenance
No conversion gasses, heating elements, quartz tubes, or columns.

Total sulfur analysis
From ultra-low sulfur diesel and gasoline to heavy fuel oil and crudes, the analyzer delivers unprecedented precision and accuracy.

For use in multiple application areas
Refinery labs, pipeline terminals, additive plants, and inspection laboratories.

Hassle-free sample prep
Utilizes Accucell pre-assembled and pre-vented sample cups for enhanced precision, extreme ease-of-use, and enhanced productivity.

Calibration curves: 8 calibration curves. Automatic and Manual Calibration functionality.
Catalysts: No
Communication Capabilities: Printer, Ethernet: LIMS Connectivity (optional)
Concentration: In ppm: 4, 8, 15, 100, 500
Consumable/Utility Gas: Vacuum
Dimensions: Depth: 19.7 in / 50.0 cm
Dimensions: Height: 13.4 in / 34.0 cm
Dimensions: Width: 14.6 in / 37.0 cm
Element: Sulfur / 0.4 ppm
Measurement Time: User selectable 30-900 s, 300 s typical
Method Name: ASTM D2622, ASTM D7039,
ISO 20884
Options: LIMS Connectivity
Range: 0.4 ppm – 10 wt%
Repeatability: ppm – r
4 – 0.4
8 – 0.7
15 – 0.9
100 – 3
500 – 6
Reproducibility: ppm – R
4 – 1.0
8 – 1.2
15 – 1.7
100 – 6
500 – 12
Sample Cup Type: Accucell
Sample Matrix: Hydrocarbon
Specific Technology: MWDXRF
Type: Benchtop
Weight:73 lbs / 32 kg
What’s included?: Quick-Start Guide, Operations Manual, radiation survey, spare fan filter, a printer with ink, printer cable, 1 roll printer paper, system fuse, 100 accucell cups, 100 mylar sample films, sample stand, spare primary window, and 2 sets calibration standards; 1 for 7039 and 1 for 2622 (6 x 10 mL each).
XRF Film Type: Mylar

Accessories, Spares, Calibration, and Training