AFIDA
Technology Guide
Direct comparison to Cetane Engine (ASTM D613)
- IP 617; EN 17155; ASTM D8183 (AFIDA) contains a statement reporting no bias to ASTM D613
- Calibrated to the original cetane number scale
- Multi point calibration using 7 blends of PRF (Primary Reference Fuels)
Conformity to ASTM D975-18a
- ASTM D8183 now included, section 5.1.10
Conformity to EN590
- Section 5.7.4 is key – approves alternative technologies
- IP 617; EN 17155; ASTM D8183 fully meet EN ISO 4259 requirements
- IP 617; EN 17155; ASTM D8183 are approved alternative test methods
Significant difference with other CVCC instruments
- These use older CVCC technology reporting DCN, Derived Cetane Number
- DCN requires a correlation formulae to arrive at test result
- These DCN instruments require considerable ‘tuning’
Key strengths of AFIDA
- Minimises technician time
- Reports Indicated CetaneNumber (ICN) – not DCN
- Calibrated using PRF’s
- Fixed test temperature – no ‘tuning’ is needed
- Direct reading – no ‘derived’ formula applied
- Very small sample volume – just 40 ml
- Fully automated with 36 place sample carousel
Technology comparisons
AFIDA ASTM D8183 | ASTM D6890 | ASTM D7170 | ASTM D7668 | D613 ENGINE | |
---|---|---|---|---|---|
Test method | ASTM D8183 IP 617 EN 17155 | ASTM D6890 EN 15195 IP 498 | ASTM D7170 EN 16144 IP 567 | ASTM D7668 EN 16715 IP 615 | ASTM D613 IP 41 |
Result | ICN | DCN | DCN | DCN | CN |
Calibration | 7 PRF blends Auto using the Cetane scale | Heptane & MCH Manual | Heptane & MCH Manual | 1 PRF & MCH Manual/Auto | PRF |
Specialist gas/water | Not required | Yes | Yes | Yes | Yes |
Temperature tuning | Not required | By operator | By operator | Automatic | n/a |
Sample volume | 40 ml | 200 ml + flush | 350 ml | 370 ml | 1 liter + flush |
User operational comparisons
AFIDA ASTM D8183 | ASTM D6890 | ASTM D7170 | ASTM D7668 | D613 ENGINE | |
---|---|---|---|---|---|
Cost of instrument | Low | High | Low | Low | Very high |
Operating complexity | Low | High | High | Medium | Very high |
Test duration | 25 mins | 20+ mins | 20+ mins | 20-30 mins | 45+ mins |
Sample handling | Automatic 36 place carousel | Manual introduction | Manual introduction | Manual introduction | Manual introduction |