16260 optional

    • Aquisys Software

      Aquisys Software – 17900

      Intelligent Laboratory Software

      SCHMIDT + HAENSCH offers the fully 21 CFR Part 11 compliant software Aquisys 3 in combination with the VariFamily instruments. The software enables you to control your polarimeters, refractometers, and density meters from anywhere. Aquisys 3 allows you to control, monitor, archive, and retrieve your measured data from anywhere. The app is fully connectable to intranet, Google Chrome, and all mobile devices.

    • Flow-through tube - 09212 & 09213

      Flow-through tube – 09212 & 09213

      Tubes with funnel and riser for Polarimeter and Coloromat®

      • Flow-through tube with funnel and riser
      • Only for Unipol, Polartronic V, Saccharomat models with plug-in connection inside of the sample compartment

      Polarimeter tubes of SCHMIDT + HAENSCH are designed for different applications. All tubes meet the ICUMSA recommendations, class A (±0.01%) and the corresponding specifications of the O.I.M.L. and the Australian
      Standard K 157 regarding construction and precision. In general: Longer measuring tubes are more reliable than short ones. Sample heterogeneity and residues from previous samples have a larger influence on the accuracy of the
      measurement when applying shorter tubes. Larger sample volumes ensure more stability. An exception is the case of
      dark samples; a shorter measuring tube might be necessary due to the high absorption of light.

    • Flow-through tube thermostatable - 09216 & 09217

      Flow-through tube thermostatable – 09216 & 09217

      Thermostatable flow-through tubewith funnel and riser
      Only for Unipol, Polartronic V, Saccharomat models with plug-in connection inside of the sample compartment
      Polarimeter tubes of SCHMIDT + HANESCH are designed for different applications. All tubes meet the ICUMSA recommendations, class A (±0.01%) and the corresponding specifications of the O.I.M.L. and the Australian
      Standard K 157 regarding construction and precision. In general: Longer measuring tubes are more reliable than short ones. Sample heterogeneity and residues from previous samples have a larger influence on the accuracy of the
      measurement when applying shorter tubes. Larger sample volumes ensure more stability. An exception is the case of
      dark samples; a shorter measuring tube might be necessary due to the high absorption of light.

    • Flow-through tube with integrated temperature sensor - 03820 & 03830

      Flow-through tube with integrated temperature sensor – 03820 & 03830

      Tubes with funnel and riser for Polarimeter and Coloromat®

      Flow-through tubes with funnel and riser are available with integrated:

      • Temperature sensor
      • Thermostats and additional hose connection for water circulator
      • Tube for manual sample changing, not thermostable
      • Only for Unipol, Polartronic V and H series, Saccharomat models with plug-in connection inside of the sample compartment

      Polarimeter tubes of SCHMIDT + HAENSCH are designed for different applications. All tubes meet the ICUMSA recommendations, class A (±0.01%) and the corresponding specifications of the O.I.M.L. and the Australian Standard K 157 regarding construction and precision. In general: Longer measuring tubes are more reliable than short ones. Sample heterogeneity and residues from previous samples have a larger influence on the accuracy of the measurement when applying shorter tubes. Larger sample volumes ensure more stability. An exception is the case of dark samples; a shorter measuring tube might be necessary due to the high absorption of light.

    • Quartz control plates

      Quartz control plates

      Calibration for Polarimeters

      Quartz control plates are used as measuring standards, to test and calibrate polarimeters.

      The surface of the thin quartz plates are polished with highest precision. For reasons of mechanical stability, plates with a low angle of rotation (between -24 °Z and +24 °Z) are produced as double quartz plates (combination of a dextro and a levorotatory plate).

      For each SCHMIDT + HAENSCH quartz control plate a factory certificate is supplied, indicating the absolute angle of rotation in degrees (°) and wavelength, i.e. 405, 546, 587, 589.3, 589.44, 633 and 882 nm.

      The angle of rotation of the quartz plate is directly correlated to its thickness and will be indicated at a temperature of 20 °C for the wavelength 589 nm.

    • T-CELL

      T-CELL

      Polarimeter tubes

      Our polarimeter tubes with integrated Peltier-elements are controlled by a separate, user-friendly electronic unit which has its own power supply.

      Temperatures are set directly via the polarimeter in special software menu.

      Even though the reference temperature for polarimetric measurement is 20 °C or 25 °C SCHMIDT+HAENSCH ’s T-cells can quickly heat and cool samples over a wider range (18 °C to 40 °C) for special applications. This versatility enables an analysis of temperature dependencies for individual substances.