Setaflash Series 3e ActiveCool Flash Point Tester – 33220-0

Setaflash Series 3e ActiveCool provides fast and reliable flash point results and requires minimum operator expertise. It has been designed to rapidly and efficiently carry out Flash/No Flash tests to determine the flash point of liquids and semi-solids in the 10 to 135 °C temperature range.

• Compliance
• Safety
• Product Quality
• 2 ml of sample and a result in less than 2 minutes

Designed for precise and accurate tests, barometric pressure is measured by an internal barometer, and correction is automatically applied to the test result. Unknown samples can be tested using ramp mode.

Setaflash Series 3 ActiveCool has a very simple user interface with a color digital display and touchscreen icons that guide the operator through a straightforward testing process. For easy record keeping, the instrument stores test results, which can be saved and transferred via a USB port.

The versatile range of Setaflash instruments is ideal for use in the laboratory, production line, or for portable test applications.

Methods: ASTM D1655, ASTM D3278, ASTM D3828, ASTM D7236, ASTM D8174, ASTM E502, IP 523, IP 524, IP 534, BS EN ISO 3679 & BS 2000-523, ISO 3680 (OBS)

Features

  • Flash/No Flash & Ramp modes
  • Electric ignitor
  • 10 to 135°C temperature range
  • Flash point test in less than 2 minutes
  • Small sample size (2 or 4ml)
  • Automatic flash detection
  • Electronic Peltier cooling
  • Inbuilt touch screen
  • Inbuilt USB port
  • Portable operation
  • °C or °F display

Rapid Results in just 1 or 2 Minutes
The Setaflash Series 3 ActiveCool is an easy-to-use instrument that can complete a flash/no flash test in less than two minutes and, in many cases, in 60 seconds or determine the flash point of a sample within a temperature range of 0 °C to 135 °C. The sample cup is heated/cooled by Peltier cells. This gives very accurate and stable sub-ambient test temperatures and the benefit of rapid temperature normalization between tests.

Ease of Operation
All functions on the Series 3 ActiveCool, including test parameters, instrument status, and test results, are accessed via the touch screen display and push button. An audible prompt is sounded when a flash is achieved. An automatic flash detector reduces the chance of misinterpretation of the test result by the operator and enhances repeatability.

Innovative Design
The Series 3 ActiveCool Flash Point Tester is lightweight and suitable for mobile use when operated from the optional 12-volt power supply adaptor.
Supplied with: 2 ml syringe, Allen key, mains lead, quick start guide, and instruction manual.

TEMPERATURE RANGE 10 to 135 °C
SAMPLE CUP MATERIAL Aluminium
TEST MODES Rapid Equilibrium and Ramp
SAMPLE SIZE 2 ml for flash points up to 100 °C 4 ml for flash points above 100 °C
TEST DURATION 1 minute below 100 °C, 2 minutes above 100 °C, or user determination 1 to 99 minutes
RAMP RATE 2 °C/min ramp
HEATING AND COOLING SYSTEM Peltier Cell
POWER 200 W (maximum)
VOLTAGE 100/240 V, 50/60 Hz
IGNITOR Electric
SIZE (HXWXD) 19.5 cm x 29.5 cm x 14 cm
WEIGHT 3 kg
CCCN CODE Tariff 90268020