AvCount Air Particle, Number SA1400-0

AvCount Air provides reliable results for determining the particle concentration in air, for applications such as clean rooms, laboratories and the monitoring of air cleanliness in hospitals.

• Simple operation
• 3 measuring channels
• 6 particle size classes
• Programmable via PC

With a simple test procedure and user friendly features, operator skill or expertise is kept to a minimum.

AvCount Air can hold up to 3 test methods with custom methods created on a PC before being uploaded.

Test progress and results for up to 6 size bands are displayed as cumulative particles/ml.

Methods: ISO 14644, ISO 21501

Results Analysis The AvCount Air can be controlled from a personal computer via the USB port. This allows the configuration of test parameters, test control and download of data. It also allows operation of a single device or multiple instruments simultaneously. The measured data is displayed and can be retrieved from the data memory later. An Excel template is available for exporting the data.

Calibration & Verification Calibration can be carried out either using NIST traceable standard solutions or by interfacing with a primary calibrated ‘Master’ AvCount Air. Calibration can be carried out at a user site or at our authorised regional laboratories.

  • Cumulative Particles/ml
  • 0.21µm to 10µm
  • Simple operation
  • High visibility screen
  • Stand-alone or PC controlled
  • Programmable via PC
Particle Size Range: 0.21µm to 10µm
Number of Measuring Channels: 3 (16 via USB)
Particle Concentration: 250,000 P/sec
Flow Rate: 28.3 l/min
Ambient Temperature Range: 5 to 40°C
Air Temperature Range: 0 to 70°C
Results: 600 internal, unlimited via USB
Software: Protrend via USB
Connectivity: Via ethernet
Voltage: 240 Vac +/-10%, 50Hz, 1A
Size (HxWxD): 150 x 215 x 260cm
Weight: 5kg
Calibration: ISO 21501 – NIST traceable

Lazar Scientific is your reliable source for Fuel Cleanliness Testing Equipment.