Setaflash Series 8 ActiveCool Flash Point Tester, 82100-0

The Setaflash Series 8 ‘ActiveCool’ is an automated closed cup flash point tester with a temperature range of -20°C to 130°C (-4°F to 266°F).

There are two modes of operation; Flash / No Flash and Ramp. The ramp mode increases the sample temperature at a rate of 2°C/min and automatically dips the ignitor every 1°C up to 100°C (every 2°C above 100°C), until either a flash is detected or the end of test temperature is reached.

This instrument operates with electric ignition.

Methods: ASTM D3278, ASTM D3828, ASTM D7236, ASTM E502, BS 2000-523, CLP Regulations EC No 1272/2008, EPA 1020 A & B, IP 523, IP 524 (Obs), IP 534, ISO 3679, ISO 3680 (Obs), UN Class 3 Non-viscous Flammable Liquids

The automatic flash point detection system is tuned to ignore halo and other false flash effects, enhancing precision and accuracy. Barometric pressure is measured by an internal barometer, and the correction is automatically applied to the test result.

Peltier cell technology is used for heating and cooling, ensuring precise temperature control and rapid cooldown after a test is completed. The ignitor uses a new electrical hot wire (for the gas option see 82110-0). The bright colour LCD display shows test parameters and the flash point test result. The text and colour of the top line indicates instrument status. All functions are accessed via menus, navigated by the keypad. Up to 64 test results can be stored in memory, each tagged with a sample reference and operator name. Results can be downloaded to a PC or printer via the RS232C port. The firmware is upgradeable for future developments.

SUPPLIED WITH: Ignitor, 2ml syringe, mains lead and instruction manual.

  • -20°C to 130°C (-4°F to 266°F)
  • Small sample size, 2 or 4ml
  • Flash/No Flash & Ramp modes
  • Electric ignitor
  • Automatic dipping and flash detection
  • ActiveCool electronic Peltier cooling
  • 64 Test memory & RS232 interface
  • °C or °F temperature display
Temperature Range: Air: 10 to 130°C (50 to 266°F) Water Assisted: -20 to 130°C (-4 to 266°F) using 99300-2
Sample Size: 2 or 4ml according to method
Ignitor: Electrical Hot Wire (For Gas option see 82110-0)
Test duration, Rapid Equilibrium Mode: 1 minute below 100°C, 2 minutes above 100°C, FAME 1 minute or user defined 1 to 99 minutes
Test duration, Ramp Mode: Typically 7 minutes
Cool Down Time 80 to 20°C (176 to 68°F) in 4 minutes
Sample Cup Material Aluminium
Heating / Cooling Method Peltier Cell
Computer Interface RS232C (USB & LAN options)
Voltage 100 to 250V, 50/60Hz (2x AA dry cell batteries required for internal clock)
Power 300W (maximum)
Size (HxWxD) 30 x 34 x 38cm
Weight 8kg
CCCN Code: Tariff 90278099