PM-93 Pensky-Martens Flash Point Tester – 35000-0
The Seta PM-93 is a fully automated Pensky-Martens closed cup flash point tester which combines strict method conformance with the latest control technology and safety systems to provide the ultimate level of fast, accurate and safe analysis.
You can determine flash points up to 400 °C using either standard mode for where the flashpoint is known or search mode for when it is not. Once complete, the PM-93 graphically displays results on the LCD screen. You can save results to the internal memory, email the results, export results to a USB, or transfer results via QR code.
SQC software analyses results calculate mean, standard deviation, standard error and repeatability and plots results in accordance with ASTM D6299.
Methods: ASTM D93, BS EN ISO 2719 & BS 2000-34, IP 34
Features
- Easy operation with touch screen
- Automatic flash point determination
- Gas or electric ignition
- Patented ‘SafeFlash’ auto fire extinguisher
- Large capacity memory for storage of 30 test profiles, operator names, and test methods
- Real-time display of test progress and storage for 2000 results
- LIMS connectivity, results can be emailed, saved to USB or transferred via QR code
- Statistical Quality Control (SQC) software to analyze results to conform to ASTM D6299
Operator safety:
Patented ‘SafeFlash’ fire extinguisher automatically activates in the event of excess flame or fire around test cup
PRT checks, safety dip before starting, over-temperature cut out, gas flame relight, gas shut off and emergency stop
Complies with ISO 2719 annex b
Ease of use:
Single action lifting arm operation
Intuitive user menu with color touch screen
A fully automatic test procedure, little operator intervention required
Enhanced test throughput:
Unique test profiles to help improve efficiency when testing a large number of samples
Fast warm-up and high capacity air cooling minimizes cycle time
Precision and accuracy:
Automatic flash point ensures test repeatability
Results are corrected for variations in atmospheric pressure using the integral barometer
Unique alignment design ensures lid and cup locate perfectly
Password protected ‘calibration’ mode allows verification of instrument
SQC enhanced software to ensure precision and accuracy
IGNITION SYSTEM | Electric or gas |
FLASH DETECTION | Thermal |
COOLING | Forced air (integral fan) |
HEATING RATE | Fast heating rate (>10 °C/min) and standard; 5.5 °C/min; 3 °C/min; 1.3 °C/min; 1 °C/min |
TEMPERATURE RANGE | Ambient +5 °C to 400 °C |
INTERNAL MEMORY | 30 programmable test profiles, operator names, test methods and, unlimited tests results |
FIRE DETECTION | Thermal with optional integral Fire Extinguisher |
SIZE/WEIGHT | 38.5 x 24 x 50 cm / 25 kg |
VOLTAGE | 230 Vac +/-15% or 110 Vac +/-15%. Frequency 50/60 Hz. Auto-sensing – no selector switch required |
CCCN CODE | Tariff 9026802 |
Accessories, Spares, Calibration, and Training
-
Pelican™ Storm Case™ for the Seta PM93 – CASE-PM93
-
Spirit Level – 23680-0
-
SafeFlash Fire Extinguisher System – 35002-0
-
PM-93 Barcode Scanner – 35003-0
-
Data Printer – 81002-3
-
Silicone Rubber Tube 4 meters – 13730-005
-
PM-93 Sample Cup – 35000-005
-
Ignitor Assembly – 35000-004
-
Refill Canister for SafeFlash Fire Extinguisher – 35000-003
-
PRT Probe – 35000-002
-
Flash Detector – 35000-001
-
Heat Sink Compound – 13740-017
-
MTVM Gas Oil – 500ml – 99851-0
-
MTVM Lubricating Oil – 500ml – 99853-2
-
MTVM Fuel Oil – 500ml – 99852-0
-
Multi Test CRM – Gas Oil 500ml – SP9000-0